Description This item was bought in error and has never been used in anger for any project.  It is perfectly functional and has no damage but there are couple of witness marks where it has been slide in and out of the housing.  There are image in the lot showing the device working by using the NI MAX software and an NI USB-9162.

NI-9211 Specifications

Definitions

Warranted  specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics  describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical  specifications describe the performance met by a majority of models.
  • Nominal  specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are  Typical  unless otherwise noted.

Related information:
  • Software Support for CompactRIO, CompactDAQ, Single-Board RIO, R Series, and EtherCAT

Conditions

Specifications are valid for the range  -40 °C   to  70 °C   unless otherwise noted.

Accuracy within typical use can vary based on chassis, mounting parameters, other modules present in the system, and installed accessories.

Input Characteristics

Number of channels

4 thermocouple channels, 1 internal autozero channel, 1 internal cold-junction compensation channel

ADC resolution

24 bits

Type of ADC

Delta-Sigma

Sampling mode

Scanned

Voltage measurement range

±80 mV

Temperature measurement ranges

Works over temperature ranges defined by NIST (J, K, T, E, N, B, R, S thermocouple types)

Conversion time

70 ms   per channel;  420 ms   total for all channels including the autozero and cold-junction channels

Common-mode voltage range

Channel-to-COM

±1.5 V

COM-to-earth ground

±250 V

Common-mode rejection ratio (0 Hz   to  60 Hz )

Channel-to-COM

95 dB

COM-to-earth ground

>170 dB

Input bandwidth (-3 dB )

15 Hz

Noise rejection (at  50 Hz   and  60 Hz )

85 dB   minimum

Overvoltage protection

±30 V   between any input and COM

Differential input impedance

20 MΩ

Input current

50 nA

Input noise

1 μV RMS

Gain error (at  -40 °C   to  70 °C )

0.06%   typical,  0.1%   maximum

Offset error (with autozero channel on)

±15 μV   typical,  ±20 μV   maximum

Gain error from source impedance

Add  0.05 ppm   per Ω when source impedance  >50 Ω

Offset error from source impedance

Add  ±0.05 μV   typical,  ±0.07 μV   maximum per Ω when source impedance  >50 Ω

Cold-junction compensation sensor accuracy

0 °C   to  70 °C

±0.6 °C   typical,  ±1.3 °C   maximum

-40 °C   to  70 °C

±1.7 °C   maximum

MTBF

633,012 hours at  25 °C ; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method

Temperature Measurement Accuracy
Measurement sensitivity[1]
With autozero channel on

Types J, K, T, E, N

<0.07 °C

Type B

<0.25 °C

Types R, S

<0.60 °C

With autozero channel off

Types J, K, T, E, N

<0.05 °C

Type B

<0.20 °C

Types R, S

<0.45 °C

The following figures show the typical and maximum errors for each thermocouple type when used with the  NI-9211   over the full temperature range and autozero on. The figures account for gain errors, offset errors, differential and integral nonlinearity, quantization errors, noise errors, and isothermal errors. The figures do not account for the accuracy of the thermocouple itself.

Figure 1.  Thermocouple Type J and N Errors

Figure 2.  Thermocouple Type K Errors

Figure 3.  Thermocouple Type T and E Errors

Figure 4.  Thermocouple Type B Errors

Figure 5.  Thermocouple Type R and S Errors

Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-COM

±30 V   maximum

Isolation

Channel-to-channel

None

Channel-to-earth ground

Continuous

250 V RMS ,  Measurement Category II

Withstand

2,300 V RMS , verified by a  5 s   dielectric withstand test

Measurement Category II Caution Do not connect the product to signals or use for measurements within Measurement  Categories III   or IV. Attention Ne pas connecter le produit à des signaux dans les catégories de mesure III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

Measurement  Category II   is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example,  115 V   for U.S. or  230 V   for Europe.

Environmental Characteristics
Temperature

Operating

-40 °C to 70 °C

Storage

-40 °C to 85 °C

Humidity

Operating

10% RH to 90% RH, noncondensing

Storage

5% RH to 95% RH, noncondensing

Ingress protection

IP40

Pollution Degree

2

Maximum altitude

2,000 m

Shock and Vibration
Operating vibration

Random

5 g RMS ,  10 Hz   to  500 Hz

Sinusoidal

5 g ,  10 Hz   to  500 Hz

Operating shock

30 g ,  11 ms   half sine;  50 g ,  3 ms   half sine;  18 shocks   at  6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Power Requirements
Power consumption from chassis

Active mode

170 mW maximum

Sleep mode

4 mW maximum

Thermal dissipation (at 70 °C)

Active mode

170 mW maximum

Sleep mode

4 mW maximum

Physical Characteristics
Screw-terminal wiring

Gauge

0.25 mm2   to  2.5 mm2   (24 AWG   to  12 AWG )   copper conductor wire

Wire strip length

7 mm   (0.28 in. )   of insulation stripped from the end

Temperature rating

90 °C , minimum

Torque for screw terminals

0.3 N · m  (2.66 lb · in. )

Wires per screw terminal

One wire per screw terminal

Connector securement

Securement type

Screw flanges provided

Torque for screw flanges

0.2 N · m   (1.80 lb · in. )

Dimensions

Visit  [link removed by eBay]   and search by module number.

Weight

150 g (5.3 oz)

  • Condition: Used
  • Brand: NATIONAL INSTRUMENTS
  • Number of Channels: 4
  • MPN: NI9211
  • DAQ Module Type: Temperature
  • Unit Quantity: 1

PicClick Insights - National Instruments NI 9211 cDAQ Temperature/Thermocouple Module, Working, VGC PicClick Exclusive

  •  Popularity - 1 watcher, 0.1 new watchers per day, 10 days for sale on eBay. Normal amount watching. 0 sold, 1 available.
  •  Best Price -
  •  Seller - 136+ items sold. 0% negative feedback. Great seller with very good positive feedback and over 50 ratings.

People Also Loved PicClick Exclusive